The Light That Learned to Read Our Blood: The Remarkable History and Future of Pulse Oximetry

The History of Medical Technology | Oxygen Monitoring

From nineteenth-century spectroscopy and wartime aviation to a Japanese engineer's brilliant discovery, fingertip oxygen monitors, COVID-19, hidden measurement bias and the next frontier of wearable medicine.


A tiny red light with an extraordinary history

A nurse clips a small plastic device onto a patient's finger. A red light glows through the skin. Within seconds, the display shows two numbers: perhaps 97% and 76 beats per minute.

The first number estimates how much of the haemoglobin in the pulsating arterial blood is carrying oxygen. The second represents pulse rate. Neither requires a needle, a blood sample or an expensive imaging system.

Today, this scene is so familiar that the technology seems almost obvious. During the COVID-19 pandemic, small fingertip pulse oximeters even became household objects, appearing alongside thermometers and blood-pressure monitors.

Yet a fingertip pulse oximeter is an extraordinary piece of engineering. It must send light through a mixture of skin, connective tissue, bone and blood; detect tiny changes caused by the heartbeat; distinguish oxygenated from deoxygenated haemoglobin; and convert imperfect optical signals into a clinically useful estimate.

Its modern form emerged from a surprising combination of discoveries in chemistry, spectroscopy, wartime aviation and Japanese biomedical engineering. The crucial breakthrough was not merely the use of light. It was learning how to separate the signal from pulsating arterial blood from everything else in the finger.

1. Why measuring oxygen is so important

Our cells require oxygen for aerobic metabolism. Oxygen diffuses from the alveoli in the lungs into pulmonary capillary blood, where most of it binds reversibly to haemoglobin inside red blood cells. The circulation then carries this oxygen to the tissues.

A small fraction of oxygen is dissolved directly in plasma, but the overwhelming majority of the blood's oxygen content is carried by haemoglobin. This gives rise to three related but distinct quantities:

  • PaO2: the partial pressure of oxygen dissolved in arterial blood, usually obtained through arterial blood-gas analysis.
  • SaO2: arterial haemoglobin oxygen saturation, measured or calculated by appropriate laboratory methods depending on the analyser.
  • SpO2: the non-invasive estimate of arterial oxygen saturation produced by a pulse oximeter.

The distinction matters. A fingertip device does not directly measure PaO2. Nor does it measure the total quantity of oxygen circulating through the body.

One useful approximation for arterial oxygen content is:

CaO2 ≈ 1.34 × Hb × SaO2 + 0.003 × PaO2

Here, haemoglobin concentration (Hb) is in g/dL, SaO2 is expressed as a fraction, PaO2 is in mmHg, and oxygen content is approximately in mL O2 per dL of blood.

Imagine two people with identical oxygen saturations of 98%. One has a haemoglobin concentration of 15 g/dL and the other has severe anaemia with 7 g/dL. Their displayed saturation may be nearly identical, yet their arterial oxygen content can be very different. A reassuring SpO2 does not rule out anaemia or inadequate tissue oxygen delivery.

2. The first breakthrough: blood changes colour when it carries oxygen

The technology owes an enormous debt to spectroscopy: the study of how matter absorbs and transmits light at different wavelengths.

In the nineteenth century, investigators showed that oxygenated and deoxygenated blood have different optical properties. Felix Hoppe-Seyler and other early researchers helped establish the relationship between haemoglobin, oxygen and characteristic absorption spectra. By the late nineteenth century, optical studies of blood had become an important part of physiological chemistry.

Oxygenated blood looks bright red; deoxygenated blood is darker red. But a pulse oximeter does not simply inspect the colour of a finger. Skin pigmentation, tissue thickness, venous blood and other structures complicate the picture.

The challenge was to measure the oxygen-dependent optical change despite these overlapping effects. Early devices attacked the problem using multiple colours of light and carefully controlled measurement conditions.

3. Beer and Lambert: the mathematical foundation

Much of absorption spectroscopy is described by the Beer–Lambert law. In an ideal homogeneous sample:

A = log10(I0/I) = εcL

Here, A is absorbance, I0 is incident light intensity, I is transmitted intensity, ε is molar absorptivity, c is concentration and L is optical path length.

With two absorbing forms of haemoglobin, the idealised absorbance contains contributions from both oxyhaemoglobin and deoxyhaemoglobin. If their wavelength-dependent absorption coefficients are known, multiple optical measurements can reveal the relative mixture.

However, a human finger is not a transparent cuvette. Light scatters repeatedly through heterogeneous tissues, so a simple Beer–Lambert calculation is not enough to produce accurate clinical SpO2. Real devices require optical design, signal processing and empirical calibration against reference blood measurements.

4. The 1930s: from optical experiments to early oximeters

In the 1930s, researchers began creating devices intended to follow oxygenation through living tissue. German physiologist Karl Matthes is especially important. In 1935, he described an ear oximeter using two optical wavelength regions, recognising the value of a second measurement to compensate for changes in blood volume and other effects.

These early instruments were ingenious but cumbersome. They struggled to separate arterial blood from venous blood and surrounding tissue. Motion, positioning and changing blood flow created major difficulties.

The field had learned to measure oxygen-dependent colour changes, but it had not yet developed a convenient and reliably continuous method of isolating arterial oxygen saturation.

5. Wartime aviation: the urgent need to monitor oxygen

World War II added a new practical urgency. Pilots flying at high altitudes could experience dangerous hypoxaemia when oxygen delivery was inadequate, sometimes before they recognised the seriousness of their impairment.

In 1942, American physiologist Glenn Allan Millikan developed a lightweight ear oximeter for monitoring oxygen saturation, especially relevant to aviation. He popularised the term oximeter. Later designs improved optical components and attempted to compensate for the changing volume of blood in the ear.

In the late 1940s, Earl Wood and colleagues developed approaches using pressure-induced changes in ear blood content to help establish a reference condition. It was sophisticated physiology, but the measurement was still not as simple as clipping a sensor onto a finger.

The great unsolved problem persisted: how could the device distinguish the optical contribution of arterial blood from the relatively static optical background?

6. Takuo Aoyagi: the engineer who found the signal in the noise

The decisive breakthrough came in Japan.

Takuo Aoyagi, an electrical engineer working at Nihon Kohden, was studying an optical technique for measuring cardiac output using a dye-dilution method. His apparatus monitored light passing through the ear.

The heartbeat caused small fluctuations in the optical signal. Initially, these pulsations were an inconvenience because they interfered with measurements of the injected dye.

Then Aoyagi recognised something profoundly useful.

Each arterial pulse briefly changes the volume of blood in the illuminated tissue. If one measures the tiny pulsatile change at two different wavelengths and compares it with the corresponding baseline signal, one can largely suppress the influence of non-pulsating tissue and isolate an oxygen-dependent arterial component.

What had looked like unwanted noise became the essential measurement signal.

Aoyagi conceived the principle in the early 1970s and presented the work in 1974. Nihon Kohden introduced its OLV-5100 ear oximeter in 1975. Akio Yamanishi and colleagues at Minolta were also developing closely related pulse-based optical approaches in Japan during this period. Minolta introduced an important fingertip device in the late 1970s.

The chronology matters. Aoyagi is rightly celebrated for establishing the modern pulse-oximetry principle, but its development also involved concurrent ideas, competing patents and subsequent engineering by multiple teams. The first successful clinical ecosystem required much more than a single insight.

Read the comprehensive historical review, Ninety years of pulse oximetry.

7. The remarkable idea behind the modern fingertip sensor

Today, a conventional transmissive pulse oximeter typically uses two light-emitting diodes: one red, at approximately 660 nm, and one near-infrared, at approximately 940 nm. A photodetector on the opposite side of the finger measures the transmitted light.

Reduced haemoglobin and oxyhaemoglobin absorb these two wavelengths differently. In the red region, deoxygenated haemoglobin tends to absorb more strongly; around 940 nm the relative relationship differs.

As the arterial blood volume changes with every heartbeat, transmitted light fluctuates slightly. Signal processing separates each wavelength's pulsatile component (AC) from its slowly varying background (DC).

The famous ratio of ratios is:

R = (ACred/DCred) ÷ (ACIR/DCIR)

The device then uses a calibration function to translate this ratio into an SpO2 estimate:

SpO2 = f(R)

That final relationship is not a universal equation derived solely from optical theory. Commercial pulse oximeters generally rely on empirical calibration and additional processing. Device algorithms may also examine signal quality, pulse detection and motion-related artefacts.

It is a beautiful solution to a difficult inverse problem: the fingerprint of pulsating arterial blood is extracted from an optical signal contaminated by tissue, venous blood and the environment.

8. What does the plethysmographic waveform tell us?

A pulse oximeter often displays a moving waveform called a photoplethysmogram, or PPG. This waveform reflects optical changes associated with pulsatile blood-volume dynamics at the measurement site.

From the timing of successive pulses, the device can estimate pulse rate. Some systems also provide a perfusion index, reflecting the relative amplitude of the pulsatile component compared with the non-pulsatile component.

But a PPG waveform is not the same as an arterial pressure waveform. Its shape depends on vascular physiology, sensor placement, contact pressure, tissue properties and optical processing.

This distinction becomes particularly important when wearable companies attempt to estimate blood pressure or vascular health from PPG signals.

9. The 1980s: the pulse oximeter enters the operating theatre

The next revolution was commercial and clinical.

In the late 1970s and early 1980s, companies such as Biox and Nellcor transformed pulse oximetry into a practical monitoring technology. Advances in red and infrared LEDs, photodiodes, electronic processing, sensor construction and instrument usability made continuous monitoring increasingly convenient.

Operating theatres became a natural application. Anaesthesia alters breathing, airway protection and physiological regulation. Before routine pulse oximetry, potentially dangerous oxygenation problems could be difficult to recognise promptly without blood sampling or indirect observation.

Continuous SpO2 monitoring gave clinicians an immediate warning signal when oxygenation began to deteriorate. Together with capnography and other developments in anaesthesia, it became an important component of improved patient-monitoring practice.

Pulse oximetry subsequently spread into intensive care, emergency departments, neonatal units, recovery rooms and ambulances. It became a routine monitor for many patients receiving sedation or respiratory support.

10. The major companies and their contributions

Company or lineageHistorical contribution
Nihon KohdenAoyagi's discovery of pulse oximetry and the 1975 OLV-5100 ear oximeter.
MinoltaEarly independent engineering and important fingertip pulse-oximetry systems.
Biox / OhmedaEarly commercially practical electronic devices in the US.
NellcorMajor role in clinical adoption, sensors and integrated monitoring.
MasimoSignal-extraction innovations aimed at motion and low-perfusion measurement; later multiwavelength systems.
Other manufacturersHospital-monitoring, home-care and wearable ecosystems involving firms such as Philips, GE HealthCare, Nonin and others.

These are influential historical lineages, not a current market-share ranking. Product portfolios and corporate ownership change over time.

11. Motion, cold fingers and weak pulses: a difficult engineering problem

A fingertip oximeter works best when there is a clear arterial pulsation and minimal interference. Real patients are less cooperative than laboratory test signals.

Moving the hand changes the optical path. A poorly attached sensor can shift or allow stray light to enter. Cold fingers and poor peripheral perfusion can greatly weaken the pulse-related component. Venous blood may move with motion, creating signals that complicate the assumption that all pulsations are arterial.

In the 1990s, companies including Masimo developed advanced signal-processing approaches intended to improve performance under motion and low perfusion. These technologies helped extend pulse oximetry into more difficult clinical settings, although performance always remains device- and condition-dependent.

Sensor design also diversified: fingertip clips, adhesive neonatal sensors, forehead reflectance probes, ear probes and integrated bedside monitoring systems address different needs.

12. Two wavelengths are clever, but they cannot tell us everything

Conventional two-wavelength pulse oximetry generally assumes that the principal absorbing haemoglobin species are oxyhaemoglobin and deoxyhaemoglobin. This works well enough for many clinical applications, but it has significant limitations.

Carbon monoxide poisoning: carboxyhaemoglobin can cause an ordinary two-wavelength pulse oximeter to display a misleadingly reassuring SpO2. A normal reading does not exclude carbon monoxide poisoning.

Methemoglobinemia: elevated methaemoglobin can also cause misleading estimates, often pulling readings toward the mid-80% range under certain conditions.

Anaemia: oxygen saturation can remain normal even when total oxygen-carrying capacity is low.

Low perfusion and movement: weak or corrupted pulsatile signals may produce unstable values or no reading at all.

Modern multiwavelength co-oximetry systems can estimate additional haemoglobin species under specified conditions, but their claims and limitations must be assessed for each device. Laboratory blood co-oximetry remains important when abnormal haemoglobin species are suspected.

13. COVID-19 turns the pulse oximeter into a household object

During the COVID-19 pandemic, pulse oximeters moved dramatically beyond hospitals. Many patients and families began using fingertip devices at home to track possible deterioration in respiratory illness.

The idea was attractive: a small device could potentially identify low oxygen saturation before obvious breathlessness became severe. The phrase silent hypoxaemia entered popular discussion.

But home monitoring also exposed a problem familiar to engineers: a number without context can mislead.

Cheap consumer devices varied widely in quality and validation. Incorrect use, cold fingers, movement and poor perfusion could distort readings. A person could become seriously ill despite a seemingly acceptable number, while transient artefacts could create unnecessary alarm.

Pulse oximetry became both a powerful example of accessible medical technology and a reminder that measurement requires interpretation.

14. A disturbing discovery: skin-pigmentation bias

One of the most important modern developments in pulse oximetry has been renewed attention to performance differences associated with skin pigmentation.

Melanin influences how light propagates through tissue. Differences in pigmentation can interact with sensor design, wavelength selection, calibration and processing. Device errors do not arise from pigmentation alone, and individual-device performance varies, but clinical research has shown potentially consequential disparities.

A widely discussed study published in the New England Journal of Medicine in 2020 examined cases in which pulse-oximeter readings appeared reasonably reassuring although directly measured arterial saturation was substantially lower.

In one cohort, among paired observations with SpO2 of 92–96%, the investigators found arterial SaO2 below 88% in 11.7% of observations involving Black patients and 3.6% involving White patients. These are findings for the specific study cohort, not universal probabilities for all patients or devices.

Such hidden low oxygenation is often called occult hypoxaemia. The findings prompted renewed scrutiny of clinical validation, device calibration and the representation of different skin pigmentations in performance studies.

In January 2025, the US Food and Drug Administration issued draft guidance proposing strengthened clinical and nonclinical performance evaluation, including attention to skin pigmentation. The draft status matters: proposed recommendations should not be presented as already-finalised universal legal requirements.

The next breakthrough in pulse oximetry may be not a smaller sensor, but a measurement that is equally trustworthy across the people it is intended to serve.

Research published in 2026 comparing multiple devices against existing and anticipated performance criteria underscores how difficult this challenge remains.

15. What does “98% accurate” actually mean?

Suppose a pulse oximeter displays 98%. It does not mean that the device is 98% accurate. It means the device estimates a haemoglobin oxygen saturation of 98%.

Accuracy must be evaluated by comparing SpO2 against an appropriate arterial reference across a range of oxygen saturations and participants.

One common summary metric is the root-mean-square error:

ARMS = √[(Σ(SpO2,i − SaO2,i)²)/N]

This metric combines bias and scatter across the study observations. But a small overall ARMS can still conceal clinically important subgroup differences or larger errors for individual measurements.

One practical implication is that a displayed difference of one percentage point should not automatically be interpreted as a real physiological change. Trends, signal quality, device performance and the clinical setting all matter.

16. Why saturation behaves differently from oxygen pressure

The relationship between blood oxygen partial pressure and haemoglobin saturation is nonlinear. It is commonly represented by the oxygen–haemoglobin dissociation curve, which has an approximately sigmoidal shape.

At relatively high PaO2, the curve is comparatively flat: substantial changes in PaO2 may cause only small changes in saturation. At lower PaO2, the curve becomes steeper, so modest pressure changes may produce larger saturation changes.

Temperature, pH, carbon dioxide and 2,3-BPG can shift the curve. This helps explain why SpO2 is a useful but incomplete indicator of oxygen transport and physiology.

The clinical interpretation of a reading must consider altitude, underlying disease, prescribed oxygen targets and the patient's condition. One universal numerical threshold is not appropriate for every setting.

17. The next generation: wearable and remote oxygen monitoring

Modern pulse oximetry increasingly operates outside the familiar spring-loaded finger clip.

Reflectance sensors place the light source and detector on the same side of tissue, enabling measurements at selected sites where transmission geometry is impractical. This concept supports some wrist-worn and adhesive devices.

Continuous wearable monitors seek to reduce movement artefacts, improve power efficiency and collect longitudinal data rather than isolated readings.

Telemedicine platforms can transmit oxygen trends and alerts to remote clinical teams, although alerts require careful design to avoid alarm fatigue and false reassurance.

Smartphone-linked devices can combine SpO2 estimates with pulse, activity and symptom records. Consumer wellness features should not automatically be treated as equivalent to clinically validated medical monitors.

Multispectral and multimodal sensors may incorporate additional wavelengths and other signals to improve robustness or estimate additional blood characteristics in specific validated applications.

AI-based signal analysis may help detect unusable measurements, classify artefacts and extract more information from the PPG waveform. But software cannot recover information that the sensor never reliably captured, and predictive performance must be tested across diverse people and conditions.

18. Could we build a pulse oximeter for a few hundred rupees?

Frugal science offers an intriguing educational challenge. A demonstration circuit can be assembled using red and infrared LEDs, a photodetector, an analogue front end, a microcontroller and a display or computer interface.

Students could examine how optical absorption changes with wavelength, measure pulse-related modulation on an appropriately designed optical test setup, and compare electronic filtering methods using synthetic or pre-recorded signals.

But measuring a pulse waveform is much easier than producing a validated oxygen-saturation estimate. Clinical accuracy depends on sensor geometry, optical characteristics, calibration populations, reference measurements, motion behaviour, safety and software.

Accordingly, an inexpensive student prototype should be treated as an educational photoplethysmography instrument, not an unvalidated clinical oxygen monitor. Tests involving people require suitable institutional consent, ethics and safety arrangements; student devices should not guide medical decisions.

This distinction reflects the deeper lesson of the Foldscope and other frugal-science innovations: the goal is to rethink the physical problem, not to remove the evidence needed to establish performance.

19. What the future still needs to solve

Several important problems remain open:

  • Equitable accuracy: validation and performance across the full spectrum of skin pigmentation and diverse patient populations.
  • Motion tolerance: reliable signals during daily activity and difficult monitoring conditions.
  • Low perfusion: better performance when peripheral arterial pulses are weak.
  • Calibration transparency: reproducible evidence on device performance and its limitations.
  • Clinical integration: alerts and dashboards that improve care without overwhelming staff.
  • Affordability and repairability: dependable sensors for communities with limited resources.
  • Beyond saturation: better integration with measures of haemoglobin, perfusion, respiratory status and tissue oxygen delivery.

The most useful future device may be unobtrusive, energy-efficient, inexpensive and almost invisible to the user. But its most valuable feature will remain the one that is hardest to see: a trustworthy measurement.

20. A small light, a profound scientific journey

The pulse oximeter began long before LEDs and microprocessors. Its intellectual roots lie in studies of light and colour, the chemistry of haemoglobin, the mathematics of optical absorption and the physiology of circulation.

Early investigators built unwieldy instruments to monitor oxygen changes in the ear. Wartime scientists needed practical monitors for aviators. Takuo Aoyagi recognised that the heartbeat could isolate arterial information from the surrounding tissue. Commercial innovators then transformed the principle into a reliable clinical ecosystem.

Later generations improved sensors, electronics, motion tolerance and portability. During the COVID-19 pandemic, the technology became an everyday household symbol of respiratory health. And ongoing research has revealed that even one of medicine's most successful monitoring devices still has serious limitations that must be addressed.

All of this engineering is hidden inside a tiny clip that shines red light through a finger.

The central challenge has not changed: how do we extract reliable biological information from a complicated living body without disturbing it?

In that sense, pulse oximetry is far more than an oxygen monitor. It is one of the clearest demonstrations of how physics, biology, mathematics, electronics and clinical medicine can come together to change healthcare.


References and further reading

  1. Quaresima V, Ferrari M, Scholkmann F. (2024). Ninety years of pulse oximetry: history, current status, and outlook. Full text.
  2. Severinghaus JW. (2007). Takuo Aoyagi: discovery of pulse oximetry. PubMed.
  3. Nihon Kohden. Takuo Aoyagi and the pulse oximeter. Company history.
  4. Severinghaus JW, Astrup PB. (1986). History of blood gas analysis. VI. Oximetry. PubMed.
  5. Jubran A. (1999). Pulse oximetry. Full text.
  6. Sjoding MW et al. (2020). Racial Bias in Pulse Oximetry Measurement. New England Journal of Medicine. DOI.
  7. US Food and Drug Administration. (January 2025). Pulse Oximeters for Medical Purposes: Draft Guidance. FDA guidance page.
  8. US Food and Drug Administration. Pulse Oximeter Basics. Consumer information.
  9. Blaney et al. (2025). Pulse Oximetry: 50 Years of Inventions and Discoveries in Biomedical Optics. Full text.
  10. Pulse Oximeter Performance and Skin Pigment: Comparison of 34 Oximeters Using Current and Emerging Regulatory Frameworks. (2026). PubMed.